| Customization: | Available |
|---|---|
| Classification: | Imaging Diagnostic Equipment |
| Type: | Patient Monitor |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency

Item | Qty |
Main unit | 1 |
5-lead ECG cable | 1 |
Disposable ECG electrode | 5 |
Adult SpO2 probe | 1 |
Adult NIBP cuff | 1 |
NIBP extension tube | 1 |
Temperature probe | 1 |
Power cable | 1 |
User manual | 1 |
Fuse | 2 |
ECG | |
Lead mode | 5 Leads(Ⅰ,II,Ⅲ,AVR, AVL, AVF, V) |
Gain | 2.5mm/mV 5.0mm/mV, 10mm/mV, 20mm/mV |
Heart rate | 15-300BPM (Adult); 15-350BPM (Neonatal); Resolution: 1 BPM Accuracy: ±1% Sensitivity: >200μV P-P |
ST Segment Range | -2.0mV~+2.0mV Accuracy: -0.8mV~+0.8mV; ±0.02mV or±10%, which is greater Other range: unspecified |
Sweep speed | 12.5mm/s, 25mm/s, 50mm/s |
Bandwidth | Surgery 1~20Hz Monitor 0.5~40Hz Diagnostic 0.05~130Hz |
SpO2 | |
Measuring range | 0~100% |
Resolution | 1% |
Accuracy | 70%~100%(±2%) |
PR | |
Pulse rate | 20-250BPM |
Resolution | 1BPM |
Accuracy | ±3BPM |
NIBP | |
Method | Oscillometry |
Measure mode | Manual, Auto, STAT |
Unit | mmHg, kPa |
Measure and alarm range | Adult Mode: SYS40~280mmHg DIA10~225mmHg MEAN20~240mmHg |
Pediatric Mode: SYS40~220mmHg DIA10~160mmHg MEAN20~170mmHg | |
Neonatal Mode: SYS40~135mmHg DIA10~100mmHg MEAN20~110mmHg | |
Resolution | 1mmHg |
Accuracy | ±5mmHg |
TEMP | |
Measure and alarm range | 0~50℃ |
Resolution | 0.1℃ |
Accuracy | ±1℃ |
RESP | |
Method | Impedancebetween RA-LL or RALA |
Measurement range | Adult: 10~120BrPM Neonatal/Pediatric: 7~150BrPM |
Resolution | 1BrPM |
accuracy | ±2 BrPM |
Standard Parameters | ECG, RESP, TEMP, NIBP, SpO2, PR |
Optional Parameters | Touch Screen, Recorder (Printer), Central monitoring system, Dual IBP, Etco2, 12-lead ECG,Nell-cor SpO2, SunTech NIBP |