Name | Multi-parameter Patient Monitor |
Model | NL070401001A-4 |
Place of Origin | CHINA | Certificate | CE ISO |
Standard Configuration | SpO2, NIBP ,TEMP,pulse rate | Optional Configuration | Independent SpO2 module, Nellcor SpO2 optional |
Power Source | Electricity | Material | Plastic, ABS+PC Fireproof Materials |
After-sale Service | Online technical support | Color LCD Display |
8-inch real color TFT screen |
Power consumption | 100V-240V~,1.0-0.5A | Warranty | 1 Year |
Instrument Classification | Class II | Customization | Yes |
Application Range | Adult/Pediatric/Neonatal | Battery | 11.1V24wh Li-ion battery |
Display | 8"real color TFT screen | 2h operating after full charge | |
Resolution 800*600 | 5min operating after low battery alarm | ||
One alarm indicator | Dimension and Weight | 164mm*270mm*128mm; 2.5kg | |
One working indicator | Package | 360m*250m*290m; 4kg | |
One battery charge state indicator | Data Storage | ||
Operating environment | Trend diagram/table | 720h | |
Temperature | 0~40ºC | NIBP review | 1000 group |
Humidity | 15%~80% | Wave review | 12h |
Pressure | 86.0kPa~106.0kPa | Alarm review | 200 alarm events |
Power Requirements | Quality Standard | CE, ISO 13485 | |
AC | 100~240V, 50Hz/60Hz | SFDA | Class II B |
DC | Built-in rechargeable battery | SpO2/NIBP | BF |
NIBP | Max Standard Deviation | 8mmHg | |
Method | Oscillometry | Overpressure protection | |
Measure mode | Manual, Auto, STAT | Adult | 300mmHg |
Measure interval in auto mode | 1~5/10/15/30/60/90/120/180/240/480min | Pediatric | 240mmHg |
Measure period in STAT mode | 5min | Neonatal | 150mmHg |
Pulse rate range | 40~240bpm | SpO2 | |
Measure & alarm range | Measurement range | 0~100% | |
Adult: | SYS 40~280mmHg | Resolution | 1% |
DIA 10~220mmHg | Accuracy | 70%~100%±2% | |
40%~69%±3% | |||
MEAN 20~240mmHg | Pulse Rate | ||
Pediatric | SYS 40~220mmHg | Measure/alarm range | 20~300bpm |
DIA 10~160mmHg | Resolution | 1bpm | |
MEAN 20~170mmHg | Accuracy | ±3bpm | |
Neonatal | SYS 40~135mmHg | NIBP cuff & tube | |
DIA 10~100mmHg | Standard accessories | SpO2 sensor | |
MEAN 20~110mmHg | Lithium-ion battery | ||
Static pressure measurement | Power cable | ||
Range | 0~300mmHg | Operator's manual | |
Accuracy | ±3mmHg | Optional accessories | Nellcor SpO2 |
Resolution | 1mmHg Accuracy | Suntech NIBP | |
MAX Mean Error | ±5mmHg |